|
Introduction
Minimum Requirements
I. Personnel
II. Buildings
III. Air
IV. Effluent
V. Equipment and Materials
Appendix 6
(ii) Annex 1
(ii) Annex 2
(ii) Annex 3
(ii) Annex 4
|
|
Extracted from the Report of the Thirtieth Session of the European Commission
for the Control of Foot-and-Mouth Disease, 1993.
FOREWORD
In 1985 the European Commission for the Control of Foot-and-Mouth Disease,
FAO, adopted a document entitle "Minimum Standards for Laboratories working
with FMDV in vitro and in vivo" describing a set of precautions
to be taken by foot-and-mouth disease (FMD) laboratories to avoid an escape
of virus. It was prepared at a time when the majority of countries on
continental Europe employed systematic annual prophylactic vaccination
of their cattle.
Although the above document dealt with all important aspects of FMD containment,
it has been found necessary to review it with special reference to the
need for more specific technical and general requirements as a consequence
of the recent change in Europe to a policy of non-vaccination. As a result
the present document has been prepared.
The security standards as specified herein should be considered as a minimum
requirement for FMD laboratories located in FMD-free countries with or
without systematic prophylactic vaccination. Even in countries where FMD
is present, it is important to avoid the escape of FMD virus from laboratories.
The standards in this document are therefore recommended as the minimum
for FMD laboratories, regardless of the prevailing disease situation.
Introduction
Foot-and-mouth disease is one of the most contagious diseases known and
manipulating the virus in the laboratory without adequate precautions
is a hazard. The escape of a single infectious unit of FMD virus from
a laboratory could potentially cause an outbreak.
The main sources of virus or infectious RNA (in increasing risk of hazard)
are:
1. Infected tissue cultures.
2. Infected baby mice, guinea pigs, rabbits etc.
3. Physical and chemical processing of large quantities of virus outside
closed vessels (e.g., concentration, purification, inactivation, etc.).
4. Infected pigs, cattle, sheep, goats and other susceptible animals.
Ways by which the virus or infectious RNA may escape or be carried out
from laboratories include:
Personnel
Air
Effluent and other waste
Equipment
Therefore all laboratories manipulating FMD virus must work under high
containment conditions. The safety precautions must preclude any escape
of virus and special attention must be given to:
- the prevention of illegal entry into the restricted area
- the presence of changing and showering facilities
- the responsible behaviour of personnel within and when they leave the
laboratory
- application of rules for primary containment
- the use of inactivated virus where possible
- the maintenance of negative air pressure where virus is manipulated
and decontamination of exhaust air.
- the decontamination of effluent
- the disposal of carcasses in a safe manner
- the decontamination of equipment and materials before removal from the
restricted area.
To achieve this containment a variety of technical installations and a
comprehensive set of disease security regulations are required under the
supervision of a Disease Security Officer.
The Disease Security Officer must regularly receive technical reports
about the various installations and monitor their performance. On the
basis of day-to-day records he prepares an annual report on security (incidents,
improvements, etc.) to the director.
FMD laboratories may be authorised to manipulate live FMD virus for:
DIAGNOSIS
and/or
LARGE SCALE VIRUS PRODUCTION
VACCINE TESTING INCLUDING INFECTION OF LARGE ANIMALS, and
EXPERIMENTAL INFECTION OF LARGE ANIMALS.
Minimum Requirements
I. Personnel
1. Control of access to the premises - prevention of illegal entry to
the restricted area.(Area where virus is manipulated and rooms in direct
and indirect contact therewith)
2. Personnel must be appropriately trained for the position held.
3. Entry into - and exit from - the restricted area must take place only
through changing and showering facilities. This means a complete change
from private to working clothes on entry - and another complete change
as well as a shower on exit.
4. Personnel must be regularly trained in disease security.
A code of disease security practice, including instructions for entry
into - and exit from - restricted areas, must be available for all employees
on site and for visitors. The disease security regulations must have been
read and signed by each employee at the beginning of their employment.
5. All staff members must be appropriately informed and regularly trained
in emergency evacuation procedures with special attention being given
to security requirements in cases of fire.
6. Personnel must contractually agree not to keep any animals which are
susceptible to FMD, nor reside on premises where such animals are kept
nor in the same household as other persons working with such animals and
to abide by minimum standards or quarantine, i.e., no contact with animals
susceptible to foot-and-mouth disease for at least three days. The same
applies to visitors.
Special care should be taken to ensure that visitors are instructed in
decontamination procedures and that these procedures are properly followed.
7. Regular supply of appropriate laboratory clothing for use within the
restricted area.
II. Buildings
8. General construction of buildings and their surfaces, including ducting
of the air conditioning system:
- well-maintained condition with a high standard of airtightness.
- insect, rodent and bird proof
9. Windows:
- non-opening and able to withstand operating pressures.
- shock proof in animal rooms.
10. Doors: - warning sign emblem at entrances:
ACCESS FOR AUTHORISED PERSONNEL ONLY
BIOLOGICAL HAZARD
- access restricted by locked doors
- airlocks provided with self-closing, airtight doors.
- doors fitted with windows where appropriate
11. Walls, floors, ceilings:
- appropriate surfaces easily cleaned sealed (airtight) entry of service
lines
III. Air
Ventilation systems
12. Air is removed from virus manipulating areas through a HEPA filtration
system which guarantees a negative pressure to atmosphere of at least
35 pascals (3.5 mm water) for laboratory rooms and rooms for small experimental
animals and 50 pascals (5 mm water) for large scale virus production rooms
and for large animal rooms.
13. The exhaust air from rooms where experiments in large animals are
carried out must pass through two HEPA filters in series.
14. The filter installations must allow for their testing and safe changing
in situ.
15. Manometers measuring the negative pressure in facilities and, where
appropriate, the pressure drop across filters must be installed. They
must be monitored and recorded regularly and, where appropriate, incorporate
alarms. Every effort should be made to prevent a positive pressure within
the building when either shutting off or turning on the ventilation system.
Input and extract fans can be interlocked so that the failure of an extract
fan shuts off input air and prevents reflux of contaminated air to the
exterior, e.g., by means of a flap valve in the air intake duct.
16. The laboratory power supply should be equipped with a back-up source
of electricity which starts without a delay of more than a few minutes
in the event of power failure, or the commercial power supplier must guarantee
a supply from an alternative source within a few minutes.
17. Vertical flow safety cabinets (Class 2), approved and checked, with
absolute filtration of exhaust air must be available for the handling
of FMD virus outside closed vessels.
Monitoring of ventilation systems
18. Ventilation systems should be continuously monitored to ensure proper
functions.
19. Before installation all filters must have passed an overall test for
efficiency (normally done by the manufacturer).
20. When HEPA filters are installed or replaced, an efficiency test must
be carried out to ensure proper installation and function. This testing
must be done at least once every 2 years, or when needed, e.g., when there
is a sudden change in pressure over the filters. Testing of filters should
be carried out by properly trained staff.
21. Among the methods acceptable for testing efficiency of filtration
are:
A. A smoke challenge and photometric detection system using either DOP
(dioctyl phthalate) or Shell Ondina (Annex 1 and 4).
B. A lithium flame photometer test for highly efficient filters (ref:
Hans Flyger and H.C. Rosenbaum, in American Industrial Hygiene Association
Journal, Vol. 26, 409-412, 1965 (Annex 2).
C. Electronic particle counting (Annex 3).
Change of HEPA filters
-22. Filters must be changed when the pressure difference exceeds certain
limits in accordance with the instructions given by the manufacturer,
or whenever needed.
Replacement of both pre- and absolute (HEPA) filters must be safe and
must take place in accordance with an authorised procedure. Strict precautions
must be taken to prevent spread of virus from filters or via contaminated
air. Replacement of filters from outside must take place after decontamination
in situ or in "safe change" air-handling units.
Decontamination of filters after use: wet autoclaving or incineration.
23. Filters in safety cabinets must be tested following installation and
thereafter on a regular basis using the same or equivalent methods as
used for testing of the general HEPA filtration system. When changed,
filters from safety cabinets must be autoclaved or otherwise decontaminated
before removal from the restricted area.
IV. Effluent
24. Effluent from laboratories and from facilities holding animals should
be treated in a manner which ensures that the inactivation of FMD virus
has been achieved. For this purpose heat or chemical treatment can be
used in a system which ensures that all the material is exposed to the
specific treatment. The entire effluent treatment system, including the
system for transport of effluent to the treatment unit, must comply with
the high containment conditions. There must be sufficient storage capacity
(tanks) for the storage of untreated effluent.
25. The equipment should have automatic monitoring systems to ensure proper
function. It must be ensured that the required temperature/pH is reached,
and that the installation will stop automatically when required limits
are reached, e.g., sterilisation temperature, time and maximum temperature
or pH for discharge of the effluent.
Liquids
(slurry, waste water)
26. Heat treatment:
100ºC for 1 hour or an equivalent heat effect.
27. Monitoring:
Automatic and continuous temperature/time/flow rate recording at different
stages of the process.
28. Chemical treatment:
NaOH or Na2CO3 or other alkaline treatment at pH12 for at least 10 hours.
Thorough mixing of the materials must be ensured. After treatment the
mixture must be neutralized and the pH must be checked before the effluent
is released.
29. Monitoring:
Automatic and continuous control and regulation of pH.
Solid waste
(animal carcasses, feedstuffs etc.)
30. Wet heat treatment (autoclave, 115ºC in the centre of all material
for 30 min or an equivalent heat effect, e.g., in a rendering process)
on site.
31. Incineration on site. The incinerators must comply with current safety
standards and be fitted with afterburners.
32. The system must exclude the possibility of re-contamination
33. Monitoring
The heating systems should be continuously monitored and recorded as part
of a fail/safe system. In the event of failure the system must be protected
as far as possible against the release of potentially infectious material.
V. Equipment and Materials
Laboratory fittings
34. Benches - impervious surfaces
35. Centrifuges, sonicators, homogenizers etc.:
- must be designed so as to contain aerosols.
36. Laboratory facilities and equipment must be cleaned and appropriately
disinfected at regular intervals. Cleansing and disinfection must be supervised
and recorded..
Handling of FMD virus
37. Primary containment must be given proper attention. Handling of FMD
virus outside closed vessels must be done in approved (Class 2) safety
cabinets.
38. Processing of large quantities of virus must take place in closed
systems. Proper attention should be paid to the decontamination of effluent
air from vessels and pipe work and to the decontamination of the processing
systems before opening for cleaning/maintenance/repair, etc.
39. Inoculation of animals and the keeping of infected animals must take
place within the restricted area under negative pressure. Personnel must
wear appropriate protective clothing when handling virus suspensions and
when inoculating or handling infected animals. On exit from animal rooms
protective clothes and footwear must be left inside the room or appropriately
decontaminated before being taken out.
Removal of equipment and materials
40. Before removal from restricted areas equipment must be decontaminated
according to the size and use of the equipment:
41. either by heat (wet autoclave), if possible at 115ºC for 30 min or
an equivalent heat effect.
42. or dry heat at 50ºC for 48 hours, in special situations, e.g., for
certain sensitive instruments.
43. or after surface disinfection, fumigation with formaldehyde (10 g/m3
at 70% relative humidity) for at least 10 minutes or (3 g/m3) for 24 hours
or equivalent with other aldehydes, e.g., glutaraldehyde or ethylene oxide
(0.8 g/litre for 1.5 h at 50ºC).
44. or thorough wash in an appropriate chemical disinfectant such as:
- 4% washing soda (Na2CO3)
- 0.5% caustic soda (NaOH)
- 0.2% citric acid
- 4% formaldehyde or equivalent with other aldehydes, e.g., glutaraldehyde
- or other disinfectant officially approved for the purpose.
Note: The efficiency of these chemical disinfectants is
considerably improved by the addition of non-ionic detergent
(concentration 0.005%).
45. Decontamination of clothing before removal from the restricted area
for laundry: heat (wet autoclave) at 115ºC for 30 min or equivalent heat
effect. The laundry process must involve at least a hot (80ºC) standard
detergent wash at some stage of the cycle. Laundering must be done on
site.
46. Treatment of papers before removal: 50ºC for 48 h, or an equivalent
heat effect, or, for single sheets, fumigation. Removal, of books should
be exceptional and under the control of the Disease Security Officer.
Fax or photocopy barrier systems should preferably be used.
47. All contaminated materials (including dead laboratory animals) must
be placed in sealed bags inside leak-proof containers and decontaminated
(e.g., by autoclaving) or transported for incineration or rendering on
site. The surface of these containers should be disinfected before removal
from the restricted area.
48. Before sending non-FMD biological material to another laboratory the
necessary precautions must be taken to ensure that the material does not
contain FMD virus. The recipient laboratory must be informed about the
potential risk of material coming from a laboratory manipulating FMD virus.
The recipient laboratory must further sign a statement that it is prepared
to receive the material and that it will take the necessary precautions.
Appendix 6 (ii) Annex 1
Air Filtration Systems at the Institute for Animal Health, Pirbright,
UK
At Pirbright air filtration systems have been installed in all buildings
in which viruses are used or experimental animals housed. The buildings
are maintained under negative pressure in relation to the atmosphere and
the extracted air is passed out through two HEPA, filters arranged in
series. There are now 25 such units covering all laboratories and animal
units
Filter Leak Testing
HEPA filters are tested by producing an oil/smoke aerosol by blowing (CO2)
carbon dioxide through a hot liquid plasticiser (Ondina Oil) which will
produce particles of 0.3 µm in diameter.
A high concentration of "smoke" particles are introduced into the air
trunking up-stream of the HEPA filters. The concentration of challenge
particles is determined by the forward light scattering photometer (J.M.
8000). When a photometer reading of between 4.0 and 4.5 (log scale) has
been reached this will indicate that there is a concentration of 80 -
100 µm of smoke per litre of air. This is considered to be a satisfactory
challenge as it is 1 x 104 times above the minimum sensitivity of the
photometer.
Holding the meter probe approximately 2.5 cm from the filter face on the
downstream side, the entire surface area including the gasket, is traversed
using slightly overlapping strokes at a rate of not more than 3 m/min.
A significant leak is described as 0.01% or more of the challenge concentration.
Frequency of HEPA filter testing
1. Filters must be tested when installed in the air handling unit.
2. When there is a sudden change in pressure over the filters.
3. Every six months in the large animal compounds.
4. If the air filtration system has been inoperative for more than one
week.
Appendix 6 (ii) Annex 2
Air Filtration Systems at the State Veterinary Institute for Virus
Research, Lindholm, Denmark
At Lindholm air filtration systems have been installed in buildings in
which FMD virus is handled or experimental animals are housed. Facilities
for virus/vaccine preparation and houses for experimental animals are
maintained under negative pressure in relation to atmosphere and the extracted
air is passed out through two HEPA filters arranged in series. There are
at present 16 such units and nine units with single HEPA filters.
Testing of filtration efficiency
HEPA filters are tested by producing an aerosol of lithium sulphate microcrystals
with a particle size of 0.3 µm. This is achieved by atomizing a 2% aqueous
solution of lithium sulphate monohydrate by a specially constructed apparatus
equipped with three Collison type atomizers. The aerosol is introduced
into the exhaust air channels in such a way that even distribution of
particles over the whole filter area is obtained. In the animal units,
where most of the air is recirculated, the aerosol is introduced into
the recirculation system, which in fact fills up the whole room with particles.
With the aerosol generator running air is sampled by controlled flow through
50 mm Ø cellulose nitrate membrane before and after HEPA filters. The
filters are transferred to measuring flasks, dissolved in nitric acid,
diluted, and lithium is assayed by atomic absorption spectrofotometry.
A retention of more than 99.95% of Li is considered as correct performance.
Contrary to the sodium flame test, which may be seriously influenced by
common dust particles, this lithium assay takes advantage of the fact
that lithium is a very rare metal in nature.
An inconvenience is the relatively long sampling time (1-8 h) required
to make sure that a sufficient load of particles has reached the filters
in relation to the sensitivity of the meter.
The method was recommended by the Danish Atomic Energy Commission originally
for use in monitoring exhaust air from nuclear reactors.
Frequency of HEPA, filter testing
Filters must be tested:
- when installed in the air handling unit
- when there is a sudden change in pressure over the filters
- at least once every 12 months in the large animal compounds,
- if the air filtration system has been inoperative for more than one
week.
Appendix 6 (ii) Annex 3
Air Filtration Systems at the Federal Research Centre
for Virus Diseases of Animals, Tübingen, Germany
At Tübingen, exhaust air filtration systems are in function in buildings
in which FMD virus is handled or in which experimental animals are housed.
These parts of the Research Centre are maintained under negative pressure
in relation to atmosphere. Extracted air is passed through a combination
of prefilters and HEPA filters. For the laboratory building, a two-chambers
system is used. When filters have to be replaced, air stream through one
chamber is uncoupled. After removal of the filters from the isolated side,
the chamber is disinfected. Subsequently, new filters are installed from
"outside". During this procedure the other chamber remains in function,
and air stream is not interrupted.
When filters have to be exchanged in animal houses a bypass system is
applied for the extracted air.
Filter Leak Testing
HEPA filters are tested for and against leakage efficiency with the aid
of an electronic particle counter, which detects particles of > 0.3 µm
(exclusion size). This apparatus is equipped with an aspirating unit formed
like a funnel and can be used for testing both the airtight site of the
frame and efficiency of the filter area. The detector is equipped with
an optical unit to detect passing particles, and an electronic regulation
system assuring a high signal to noise ratio. Results are continuously
shown on a display and printed out upon release.
Appendix 6 (ii) Annex 4
Air Filtration Systems at the Centraal Diergeneeskundig Instituut,
(CDI), Lelystad, The Netherlands
At the CDI high containment building in Lelystad, air filtrating systems
have been installed in all buildings in which viruses are used or experimental
animals are housed.
Different levels of negative pressure in relation to the atmosphere are
maintained in these buildings. The exhaust air passes a combination of
prefilters and HEPA-EU 13 filters. The exhaust air from the rooms that
may contain high concentrations of FMDV passes two HEPA filters arranged
in series. At present there are 272 filter units in use.
Gasket testing:
Most of the air handling units are equipped with a rill for testing the
gaskets of the filters. A filter is discarded if a leak of more than 2
1/h at a pressure of 200 Pa is detected.
Efficiency testing:
The efficiency of the filters is measured in situ by generating
a DOP-oil aerosol (ATI TDAa4A) which contains particles of 0.3 µm. A high
concentration of those particles is introduced into the air upstream of
the HEPA filter. The concentration of particles in the downstream air
is measured by a ATI TDA2E photometer. A filter is discarded if the efficiency
is less than 99.995% of the challenge concentration (equivalent to a hole
of 1 mm in diameter in a filter surface of 18 m2). The tests are performed
by a member of the technical staff in collaboration with a member of the
Safety and Quality Assurance Department.
Frequency of testing:
Filters are tested when installed in the air handling units and every
two years if they have not already been discarded. Prior to incinerating,
the filters are fumigated with formaldehyde in situ.
|